Accu-Chek Guide
Error codes E-11
The test strip may be damaged.
Retest with a new test strip. If the message reappears, please contact Accu-Chek Customer Care.
The test strip may be damaged.
Retest with a new test strip. If the message reappears, please contact Accu-Chek Customer Care.
Your blood sample may contain a high level of ascorbate.
Contact your healthcare professional.
Fluid or foreign material may be present in the test strip slot.
Remove and reinsert the test strip or repeat the blood glucose or control test with a new test strip.
If the problem persists, contact Accu-Chek Customer Care.
The blood glucose result has not been sent to the default paired device. The transfer is still pending.
Place meter and paired device closer together.
There are no test results in range for the selected data.
The batteries are almost out of power.
Change the batteries now. If the symbol reappears after the batteries have been replaced, remove the batteries again, press and hold the Power/Set/OK button for at least 2 seconds, then reinsert the batteries.
| Possible Cause: | Solution |
|---|---|
| Your batteries are dead. | Insert new batteries. |
| Display is damaged or the meter is defective. | Contact Accu-Chek Customer Care. |
| Extreme temperatures. | Move the meter to a location within the stated temperature limits in your owner's manual. |
| The meter has timed out with a USB connection. | Removed the USB cable and press the power/set button to resume testing or review test results. |
The connection between the meter and PC was lost.
Disconnect and reconnect the USB cable and retry the connection. Contact Roche if the connection is lost again.
Data could not be transferred from the meter to the PC.
Check the PC or USB cable.