Accu-Chek Guide Link
Error codes Logbook - No stored results
There are no results in the Logbook.
There are no results in the Logbook.
There is an invalid result in the Logbook.
Blood glucose may be lower than the measuring range of the system.
See the Unusual Blood Glucose Results section in the User Manual.
The date entered is not valid.
Enter the correct date.
There are no results stored in the meter for the selected data.
Blood glucose may be higher than the measuring range of the system.
See the Unusual Blood Glucose Results section in the User Manual.
A meter setting was changed while in Flight Mode.
The setting change will not take effect until Flight Mode is turned off.
Pairing to a device cannot be performed while in Flight Mode.
Retry pairing when the meter is not in Flight Mode.
The batteries may be out of power.
Turn the meter back on. If you are in a cold environment, move to a location with a more moderate temperature and retest.
If the message continues to appear after several attempts, replace the batteries.
If the message reappears after the batteries have been replaced, remove the batteries, press and hold the Power/Set/OK button for at least 2 seconds, then reinsert the batteries.
The temperature is above or below the proper range for the system.
Refer to the test strip package insert for system operating conditions. Move to an area with the appropriate conditions and repeat the blood glucose or control test. Do not artificially heat or cool the meter.